Anal Chem ; 94(38): 13171-13180, 2022 09 27.
Article in English | MEDLINE | ID: covidwho-2028625

ABSTRACT

An electrochemical platform for generating and controlling a localized pH microenvironment on demand is proposed by employing a closed-loop control algorithm based on an iridium oxide pH sensor input. We use a combination of solution-borne quinones and galvanostatic excitation on a prepatterned indium tin oxide (ITO) working electrode to modulate pH within a very well confined, small volume of solution close to the electrode surface. We demonstrate that the rate of pH change can be controlled at up to 2 pH s-1 with an excellent repeatability (±0.004). The desired pH microenvironment can be stably maintained for longer than 2 h within ±0.0012 pH. As a high-impact application of the platform technology, we propose a single-step immunoassay and demonstrate its utility in measuring C-reactive protein (CRP), a critical inflammatory marker in various conditions such as myocardial infarction and even SARS-Cov-2. Utilizing pH modulation technology along with pH-sensitive fluorescence dye simplifies the immunoassay process into a single-step, where a mixture of all of the reagents is incubated only for 1 h without any washing steps or the need to change solution. This simplified immunoassay process minimizes the hands-on time of the end-user and thus decreases technician-driven errors. Moreover, the absence of complicated liquid-handling hardware makes it more suitable and attractive for an ultracompact platform to ultimately be used in a point-of-care diagnostic assay.


Subject(s)
Biosensing Techniques , COVID-19 , C-Reactive Protein , Electrochemical Techniques , Humans , Hydrogen-Ion Concentration , Immunoassay , Quinones , SARS-CoV-2
SELECTION OF CITATIONS
SEARCH DETAIL